Main technical parameter
|
Volume(length×width
×height)
|
370mm×340mm ×1570mm
|
Weight
|
about 90kg
|
|
Power supply voltage
|
AC220V±10%
|
|
Frequency
|
50HZ ~ 60 HZ
|
|
Power
|
1500W
|
|
Blood pump/spare pump
|
Flux
|
15 ~ 340ml/min (D 6mm)
|
|
20 ~ 460 ml/min (D 8mm)
|
|
Heparin pump
|
Flux
|
0.1ml/h~10ml/h(±5%)
|
Precise
|
0.1ml/h
|
|
Injector size
|
20ml/30ml/50ml 10ml/20ml/30ml (selectable)
|
|
Arterial pressure
|
Display scope
|
-300mmHg~+300mmHg (±10mmHg)
|
Venous pressure
|
Display scope
|
-50mmHg~+300mmHg (±10mmHg)
|
Dialysate flow
|
300ml/min~800ml/min linearity adjustable(+10%)(-5%)
|
|
Temperature
|
35.0ºC~39.0ºC
|
|
Resolving rate
|
0.1ºC
|
|
Conductivity
|
13mS/cm~15.5mS/cm(±0.1 mS/cm)
|
|
UF flow scope
|
0~1800ml/h(±30ml/h ) 0~4000ml/h(optional)
|
|
ISO UF flow scope
|
0~2000ml/h(±30ml/h)
|
|
TMP
|
Scope
|
-100mmHg ~ +600mmHg (±20mmHg)
|
Blood leakage monitor
|
over 1ml blood per liter dialysate (flow:500ml/min)
|
|
Blood level monitor
|
ultrasonic sensor
|
Infrared and response threshold value
|
Single air bubble of 200u I exists when Blood flux is 200ml/min
|
Inflow pressure
|
0MPa~0.6MPa
|
Inflow temperature
|
5ºC~30ºC
|
Environment temperature
|
10ºC~30ºC, Relative humidity≤70%
|
Rinse/disinfection
|
chemical disinfection (citric acid,peracetic acid and oxalic acid)
|
Hot rinse
|
80ºC
|
Back-up power supply
|
last for 15-30min after electric-cut.
|